Fatigue Testing UHMWPe Tibial Bearing Components
Fatigue strength is directly related to the underlying thickness of ultra-high-molecular-weight polyethylene (UHMWPE) elements within the tibial component of total knee prostheses.
For any component fabricated with an overall structural thickness below 9 mm, a comprehensive fatigue testing validation report is highly recommended. The FDA specifies testing a minimum of five individual samples under varying load envelopes to develop a robust asset fatigue curve (AF/N), requiring at least one primary sample to successfully complete a 10-million-cycle runout threshold.

Recommended Test Machines
800 Series Fatigue Test Machines
Static and alternating dynamic fatigue forces up to 50 kN (11,250 lbf) paired with frequency performance ratings up to 75 Hz. Employs a completely oil-free, all-electric actuator ideal for maintaining pristine, cleanroom-compliant medical test environments while delivering lower procurement, operation, and maintenance utility footprints than hydraulic actuators.