ISO 7864 Single-Use Sterile Hypodermic Needles

ISO 7864 Single-Use Sterile Hypodermic Needles

ISO 7864 specifies the following requirements for needles of nominal outside diameters 0.3 mm and 1.2 mm: nomenclature for components, cleanliness, limits for acidity and alkalinity, size designation, color coding, needle hub, sheath, needle tube, needle point, performance, packaging, labeling, storage container, transport wrapping.

What test machine requirements are needed per ISO 7864?

  • Drag Force Test (ISO 7864 Annex D) the needle pierces a substrate at a constant test speed while measuring force peak and friction force
    • Constant speed of 100 mm/min (4 ipm)
    • Report peak force and drag force
  • Bonding strength test (ISO 7864 Annex E) measures the strength of the bond between the needle tube and needle hub
    • Maximum load 500 N (110 lb)
    • Constant speed of 50 mm/min (2 ipm)

Specialized Test Machines

Micro Test Machines | Static Universal

Static Universal

Entry-level system for strength and material characterization.
Long travel single column frame ideal for highly elastic materials.


Micro Test Machines | Static and Dynamic

Static and Dynamic

Fast dynamic actuator for static, fatigue, and dynamic tests.
Produces higher loads (up to 2200 N) for low-force applications.
Dual column dynamic frame for environmental baths and chambers.
Micro Test Machines | Multichannel

Multichannel

Axial-torsional static test machine (e.g., Luer testers).
Planar biaxial system for complex in-plane loading simulations.
Three-channel XYZ tests for complex material analysis.
Need Help?

Talk to one of our Application Engineers to help you find the right solution for your test applications

Contact Us Now